Fluke 754 Documenting Process Calibrator-HART
The 754 is everything you needed and wished for in the 744 and more! HART communication capability is combined to deliver an integrated communicating calibrator. This rugged, reliable tool is ideal for calibrating, maintaining, and troubleshooting HART and other instrumentation.
The 754 does the work of several tools - sourcing, simulating and measuring pressure, temperature, and electrical signals in one rugged, hand-held device. You¡¯ll fall in love with the new improved graphical screen, Li-on battery for longer life, USB port and new accessories to complete the package.
For documentation, the 754 automates calibration procedures and captures your data. And, of course, it helps you meet rigorous standards like ISO 9000, FDA, EPA and OSHA regulations.
Supports popular instrumentation management software. The 753 and 754, 743 and 744 work with the Fluke DPC/TRACK2¢â software, and with popular programs from Honeywell Meridium, Emerson, Cornerstone, Yokogawa, Prime Technologies, Intergraph and others. It allows you to create procedures, instructions, and action lists to deliver fast, easy documentation.
The Fluke 754 Documenting Process Calibrator-HART: Work smarter. Work faster.
The 753 and 754, 743 and 744 work with the Fluke software, and with popular programs from Honeywell Meridium, Emerson, Cornerstone, Yokogawa, Prime Technologies, Intergraph and others. It allows you to create procedures, instructions, and action lists to deliver fast, easy documentation.
The 754 is a power multifunction documenting calibrator that lets you download procedures, lists, and instructions created with software-or upload data for printing, archiving, and analysis. The 754 also features a powerful built in HART? interface capable of performing nearly all the day-to-day tasks you now perform with a separate communicator.
|
HART Capabilities
The 754 is designed to take on nearly all the day-to-day tasks you now perform with a separate communicator. In fact, it offers the communication capabilities of the 375 HART communicator.
|
Versatile HART Protocol support
Manufacturer | Pressure instruments | Temperature instruments | Coriolis instruments |
ABB/Kent-Taylor | 600T | 658T1 | |
ABB/Hartmann & Braun | Contrans P1, AS 800 Series | ||
Endress & Hauser | CERABAR S, CERABAR M, DELTABAR S | TMT 1221,TMT 1821, TMT 1621 | |
Foxboro Eckardt | TI/RTT201 | ||
Foxboro/Invensys | I/A Pressure | ||
Fuji | FCX FCXAZ | FRC | |
Honeywell | ST3000 | STT25T1, STT25H1 | |
Micro Motion | 2000 2000 IS 9701 9712 9739 | ||
Moore Products | 3441 | ||
Rosemount | 1151 2088 3001C 3051, 3051S | 3044C 644 3144 3244, 3144P | |
Siemens | SITRANS P DS SITRANS P ES | ||
SMAR | LD301 | TT3011 | |
Viatran | I/A Pressure | ||
Wika | UNITRANS | T32H1 | |
Yokogawa | EJA | YTA 110, 310 and 320 |
1Sensor Trim not supported
The 754 supports the commands contained in HART protocol Version 5.7. With 2 MB of memory, the 754 supports a substantial set of HART instructions:
|
Measurement Accuracy | |||||||||||||||||
Voltage DC |
| ||||||||||||||||
Voltage AC |
| ||||||||||||||||
Current DC |
| ||||||||||||||||
Resistance |
| ||||||||||||||||
Frequency |
|
Source Accuracy | ||||||||||||||||
Voltage DC |
| |||||||||||||||
Current DC |
| |||||||||||||||
Resistance |
| |||||||||||||||
Frequency |
|
Technical Data | |||||||||
Data Log Functions |
| ||||||||
Ramp Functions |
| ||||||||
Loop Power Function |
| ||||||||
Step Functions |
|
Environmental Specifications | |||
Operating Temperature |
| ||
Storage Temperature |
| ||
Dust/water resistance |
| ||
Operating Altitude |
|
Safety Specifications | |||
Agency Approvals |
|
Mechanical & General Specifications | |||||||||||
Size |
| ||||||||||
Weight |
| ||||||||||
Batteries |
| ||||||||||
Battery Life |
| ||||||||||
Battery Replacement |
| ||||||||||
Side port connections |
| ||||||||||
Data storage capacity |
| ||||||||||
90 day specifications |
|
Temperature, Resistance Temperature Detectors | |||||||
Degrees or % of reading | |||||||
Type (¥á) | Range ¡ÆC | Measure ¡ÆC1 | Source current | Source ¡ÆC | Allowable current2 | ||
1 year | 2 year | 1 year | 2 year | ||||
100 ¥Ø Pt (385) | -200 to 100 100 to 800 | 0.07 ¡ÆC 0.02 % + 0.05 ¡ÆC | 0.14 ¡ÆC 0.04 % + 0.10 ¡ÆC | 1 mA | 0.05 ¡ÆC 0.0125 % + 0.04 ¡ÆC | 0.10 ¡ÆC 0.025 % + 0.08 ¡ÆC | 0.1 mA to 10 mA |
200 ¥Ø Pt (385) | -200 to 100 100 to 630 | 0.07 ¡ÆC 0.02 % + 0.05 ¡ÆC | 0.14 ¡ÆC 0.04 % + 0.10 ¡ÆC | 500 ¥ìA | 0.06 ¡ÆC 0.017 % + 0.05 ¡ÆC | 0.12 ¡ÆC 0.034 % + 0.10 ¡ÆC | 0.1 mA to 1 mA |
500 ¥Ø Pt (385) | -200 to 100 100 to 630 | 0.07 ¡ÆC 0.02 % + 0.05 ¡ÆC | 0.14 ¡ÆC 0.04 % + 0.10 ¡ÆC | 250 ¥ìA | 0.06 ¡ÆC 0.017% + 0.05 ¡ÆC | 0.12 ¡ÆC 0.034 % + 0.10 ¡ÆC | 0.1 mA to 1 mA |
1000 ¥Ø Pt (385) | -200 to 100 100 to 630 | 0.07 ¡ÆC 0.02 % + 0.05 ¡ÆC | 0.14 ¡ÆC 0.04% + 0.10 ¡ÆC | 150 ¥ìA | 0.06 C 0.017 % + 0.05 ¡ÆC | 0.12 C 0.034 % + 0.10 ¡ÆC | 0.1 mA to 1 mA |
100 ¥Ø Pt (3916) | -200 to 100 100 to 630 | 0.07 ¡ÆC 0.02 % +0.05 ¡ÆC | 0.14 ¡ÆC 0.04 % +0.10 ¡ÆC | 1 mA | 0.05 ¡ÆC 0.0125 % + 0.04 ¡ÆC | 0.10 ¡ÆC 0.025 % + 0.08 ¡ÆC | 0.1 mA to 10 mA |
100 ¥Ø Pt (3926) | -200 to 100 100 to 630 | 0.08 ¡ÆC 0.02 % +0.06 ¡ÆC | 0.16 ¡ÆC 0.04 % +0.12 ¡ÆC | 1 mA | 0.05 ¡ÆC 0.0125 % + 0.04 ¡ÆC | 0.10 ¡ÆC 0.025 % + 0.08 ¡ÆC | 0.1 mA to 10 mA |
10 ¥Ø Cu (427) | -100 to 260 | 0.2 ¡ÆC | 0.4 ¡ÆC | 3 mA | 0.2 ¡ÆC | 0.4 ¡ÆC | 0.1 mA to 10 mA |
120 ¥Ø Ni (672) | -80 to 260 | 0.1 ¡ÆC | 0.2 ¡ÆC | 1 mA | 0.04 ¡ÆC | 0.08 ¡ÆC | 0.1 mA to 10 mA |
1For two and three-wire RTD measurements, add 0.4¡ÆC to the specifications.
2Supports pulsed transmitters and PLCs with pulse times as short as 1 ms
Temperature, Thermocouples | |||||
Type | Source ¡ÆC | Measure ¡ÆC | Source ¡ÆC | ||
1 year | 2 years | 1 year | 2 years | ||
E | -250 to -200 | 1.3 | 2.0 | 0.6 | 0.9 |
-200 to -100 | 0.5 | 0.8 | 0.3 | 0.4 | |
-100 to 600 | 0.3 | 0.4 | 0.3 | 0.4 | |
600 to 1000 | 0.4 | 0.6 | 0.2 | 0.3 | |
N | -200 to -100 | 1.0 | 1.5 | 0.6 | 0.9 |
-100 to 900 | 0.5 | 0.8 | 0.5 | 0.8 | |
900 to 1300 | 0.6 | 0.9 | 0.3 | 0.4 | |
J | -210 to -100 | 0.6 | 0.9 | 0.3 | 0.4 |
-100 to 800 | 0.3 | 0.4 | 0.2 | 0.3 | |
800 to 1200 | 0.5 | 0.8 | 0.3 | 0.3 | |
K | -200 to -100 | 0.7 | 1.0 | 0.4 | 0.6 |
-100 to 400 | 0.3 | 0.4 | 0.3 | 0.4 | |
400 to 1200 | 0.5 | 0.8 | 0.3 | 0.4 | |
1200 to 1372 | 0.7 | 1.0 | 0.3 | 0.4 | |
T | -250 to -200 | 1.7 | 2.5 | 0.9 | 1.4 |
-200 to 0 | 0.6 | 0.9 | 0.4 | 0.6 | |
0 to 400 | 0.3 | 0.4 | 0.3 | 0.4 | |
B | 600 to 800 | 1.3 | 2.0 | 1.0 | 1.5 |
800 to 1000 | 1.0 | 1.5 | 0.8 | 1.2 | |
1000 to 1820 | 0.9 | 1.3 | 0.8 | 1.2 | |
R | -20 to 0 | 2.3 | 2.8 | 1.2 | 1.8 |
0 to 100 | 1.5 | 2.2 | 1.1 | 1.7 | |
100 to 1767 | 1.0 | 1.5 | 0.9 | 1.4 | |
S | -20 to 0 | 2.3 | 2.8 | 1.2 | 1.8 |
0 to 200 | 1.5 | 2.1 | 1.1 | 1.7 | |
200 to 1400 | 0.9 | 1.4 | 0.9 | 1.4 | |
1400 to 1767 | 1.1 | 1.7 | 1.0 | 1.5 | |
C | 0 to 800 | 0.6 | 0.9 | 0.6 | 0.9 |
800 to 1200 | 0.8 | 1.2 | 0.7 | 1.0 | |
1200 to 1800 | 1.1 | 1.6 | 0.9 | 1.4 | |
1800 to 2316 | 2.0 | 3.0 | 1.3 | 2.0 | |
L | -200 to -100 | 0.6 | 0.9 | 0.3 | 0.4 |
-100 to 800 | 0.3 | 0.4 | 0.2 | 0.3 | |
800 to 900 | 0.5 | 0.8 | 0.2 | 0.3 | |
U | -200 to 0 | 0.6 | 0.9 | 0.4 | 0.6 |
0 to 600 | 0.3 | 0.4 | 0.3 | 0.4 | |
BP | 0 to 1000 | 1.0 | 1.5 | 0.4 | 0.6 |
1000 to 2000 | 1.6 | 2.4 | 0.6 | 0.9 | |
2000 to 2500 | 2.0 | 3.0 | 0.8 | 1.2 | |
XK | -200 to 300 | 0.2 | 0.3 | 0.2 | 0.5 |
300 to 800 | 0.4 | 0.6 | 0.3 | 0.6 |