Multifunction Process Calibrator(FLUKE/754), 0.02%
  • Á¦Á¶»çFLUKE
  • ¸ðµ¨¸í754
  • ÁÖ¿ä±â´É ´Ù±â´É±³Á¤±â

»óÇ°»ó¼¼Á¤º¸

 










 

 

 

Fluke 754 Documenting Process Calibrator-HART

The 754 is everything you needed and wished for in the 744 and more! HART communication capability is combined to deliver an integrated communicating calibrator. This rugged, reliable tool is ideal for calibrating, maintaining, and troubleshooting HART and other instrumentation.
The 754 does the work of several tools - sourcing, simulating and measuring pressure, temperature, and electrical signals in one rugged, hand-held device. You¡¯ll fall in love with the new improved graphical screen, Li-on battery for longer life, USB port and new accessories to complete the package.
For documentation, the 754 automates calibration procedures and captures your data. And, of course, it helps you meet rigorous standards like ISO 9000, FDA, EPA and OSHA regulations.
Supports popular instrumentation management software. The 753 and 754, 743 and 744 work with the Fluke DPC/TRACK2¢â software, and with popular programs from Honeywell Meridium, Emerson, Cornerstone, Yokogawa, Prime Technologies, Intergraph and others. It allows you to create procedures, instructions, and action lists to deliver fast, easy documentation.
The Fluke 754 Documenting Process Calibrator-HART: Work smarter. Work faster.

 


The 753 and 754, 743 and 744 work with the Fluke software, and with popular programs from Honeywell Meridium, Emerson, Cornerstone, Yokogawa, Prime Technologies, Intergraph and others. It allows you to create procedures, instructions, and action lists to deliver fast, easy documentation.

 

Work smarter. Work faster.

The 754 is a power multifunction documenting calibrator that lets you download procedures, lists, and instructions created with software-or upload data for printing, archiving, and analysis. The 754 also features a powerful built in HART? interface capable of performing nearly all the day-to-day tasks you now perform with a separate communicator.

  • Measure volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors, transmitters and other instruments
  • Source/simulate volts, mA, thermocouples, RTDs, frequency, ohms, and pressure to calibrate transmitters
  • Power transmitters during test using loop supply with simultaneous mA measurement
  • Measure/source pressure using any of 29 Fluke 700Pxx Pressure Modules
  • Create and run automated as-found/as-left procedures to satisfy quality programs or regulations. Record and document results
  • Holds up to a full week of downloaded procedures and calibration results.
  • Use many features like autostep, custom units, user entered values during test, one-point and two-point switch testing, square root DP flow testing, programmable measurement delay etc.
  • Easy to use
  • Three year warranty
  • Bright white dual display. Read both sourced and measured parameters simultaneously.
  • Multi lingual interface
  • Rechargeable Li-Ion battery for 10 hour uninterrupted use. Includes gas gauge.
  • Handling of fast pulsed RTD transmitters and PLCs, with pulses as short as 1 ms.
  • Delivered with DPC/Track Sample software.
  • Compatible with many Asset Management software packages.

HART Capabilities
The 754 is designed to take on nearly all the day-to-day tasks you now perform with a separate communicator. In fact, it offers the communication capabilities of the 375 HART communicator.

  • Requires no external box or second tool for everyday HART calibration and maintenance.
  • Offers fast HART communication.
  • Supports popular models of HART transmitters, with more devise-specific command support than any other HART field calibrator.
  • Works with multiple masters, burst mode, and multi-drop configurations.
  • Is easy to update as additional instruments are added and new HART versions are released.
  • Interrogate to determine device type, manufacturer, model, tag.
  • Reconfigure the sensor mapping of dual sensor temperature transmitters.
  • Read HART PV function and smart transmitter digital output while measuring analog mA output.
  • Read and write HART configuration functions to make field adjustments to PV range points, damping, and other top-level configuration settings.
  • Re-label smart transmitters by reading and writing to the HART tag field.


 

Versatile HART Protocol support

ManufacturerPressure instrumentsTemperature instrumentsCoriolis instruments
ABB/Kent-Taylor600T658T1 
ABB/Hartmann & BraunContrans P1,
AS 800 Series
  
Endress & HauserCERABAR S,
CERABAR M,
DELTABAR S
TMT 1221,TMT 1821,
TMT 162
1
 
Foxboro Eckardt TI/RTT201 
Foxboro/InvensysI/A Pressure  
FujiFCX
FCXAZ
FRC 
HoneywellST3000STT25T1,
STT25H
1
 
Micro Motion  2000
2000 IS
9701
9712
9739
Moore Products 3441 
Rosemount1151
2088
3001C
3051, 3051S
3044C
644
3144
3244, 3144P
 
SiemensSITRANS P DS
SITRANS P ES
  
SMARLD301TT3011 
ViatranI/A Pressure  
WikaUNITRANST32H1 
YokogawaEJAYTA 110, 310
and 320
 

1Sensor Trim not supported

The 754 supports the commands contained in HART protocol Version 5.7. With 2 MB of memory, the 754 supports a substantial set of HART instructions:

  • Universal commands provide functions that are implemented in all field devices, for example, read manufacturer and device type, read primary variable (PV), or read current output and percent of span.
  • Common practice commands provide functions that are common to many but not all field devices, for example read multiple variables, set damping time, or perform loop test.
  • Device-specific commands provide functions that are unique to a particular field device, for example sensor trim.

 

 

 

Measurement Accuracy
Voltage DC
  1 Year 2 Years
100.000 mV       0.02%+0.005 mV 0.03%+0.005 mV
3.00000 V 0.02%+0.00005 V 0.03%+0.00005 V
30.0000 V 0.02%+0.0005 V 0.03%+0.0005 V
300.00 V 0.05%+0.05 V 0.07%+0.05 V
Voltage AC
Range
40 to 500 Hz     
 Resolution 1 Year2 Years 
3.000 V 0.001 V 0.5%+0.002 V 1.0%+0.004 V
30.00 V 0.01 V 0.5%+0.02 V 1.0%+0.04 V
300.0 V 0.1 V 0.5%+0.2 V 1.0%+0.2 V
Current DC
  1 Year 2 Years
30.000 mA         0.01% + 5 uA  0.015%+7 uA
110.00 mA 0.01% + 20 uA 0.015%+30 uA
Resistance
  1 Year 2 Years
10.000 ¥Ø           0.05% + 50 m¥Ø 0.07%+70 m¥Ø
100.00 ¥Ø 0.05% + 50 m¥Ø 0.07%+70 m¥Ø
1.0000 k¥Ø  0.05% + 500 m¥Ø 0.07%+0.5 ¥Ø
10.000 k¥Ø 0.1% + 10 ¥Ø 0.15%+15 ¥Ø
Frequency
  Resolution Accuracy (2 Years)
1.00 to 110.00 Hz 0.01 Hz 0.05 Hz
110.1 to 1100.0 Hz 0.1 Hz 0.5 Hz
1.101 to 11.000 kHz  0.001 kHz 0.005 kHz
11.01 to 50.00 kHz 0.01 kHz 0.05 kHz

 

Source Accuracy
Voltage DC
  1 Year 2 Years
100.000 mV          0.01%+0.005 mV 0.015%+0.005 mV
1.00000 V 0.01%+0.00005 V 0.015%+0.0005 V
15.0000 V 0.01%+0.0005 V 0.015%+0.0005 V
Current DC
  1 Year2 Years 
22.000 mA (Source) 0.01%+ 0.003 mA 0.02%+ 0.003 mA
Current Sink (Simulate) 0.02% + 0.007 mA 0.04% + 0.007 mA
Resistance
  1 Year 2 Years
10.000 ¥Ø              0.01% + 10 m¥Ø 0.015% + 15 m¥Ø
100.00 ¥Ø 0.01% + 20 m¥Ø 0.015% + 30 m¥Ø
1.0000 k¥Ø 0.02% + 0.2 ¥Ø 0.03% + 0.3 ¥Ø
10.000 k¥Ø 0.02% + 3 ¥Ø 0.03% + 5 ¥Ø
Frequency
  2 Years
0.1 to 10.99 Hz 0.01 Hz
0.01 to 10.99 Hz 0.01 Hz
11.00 to 109.99 Hz 0.1 Hz
110.0 to 1099.9 Hz 0.1 Hz
1.100 to 21.999 kHz 0.002 kHz
22.000 to 50.000 kHz         0.005 kHz

 

Technical Data
Data Log Functions
Measure functions: Voltage, current, resistance, frequency, temperature, pressure
Reading rate: 1, 2, 5, 10, 20, 30, or 60 readings/minute
Maximum record length: 8000 readings (7980 for 30 or 60 readings/minute)
Ramp Functions
Source functions: Voltage, current, resistance, frequency, temperature
Rate: 4 steps/second
Trip detect: Continuity or voltage (continuity detection not available when sourcing current)
Loop Power Function
Voltage: Selectable, 26 V
Accuracy: 10%, 18 V minimum at 22 mA
Maximum current: 25 mA, short circuit protected
Maximum input voltage: 50 V DC
Step Functions
Source Functions Voltage, current, resistance, frequency, temperature
Manual Step Selectable step, change with arrow buttons
Autostep Fully programmable for function, start delay, stepvalue, time per step, repeat

 

Environmental Specifications
Operating Temperature
  -10 ¡ÆC to +50 ¡ÆC
Storage Temperature
  -20 ¡ÆC to +60 ¡ÆC
Dust/water resistance
  Meets IP52, IEC 529
Operating Altitude
  3000 m above mean sea level (9842 ft)

 

Safety Specifications
Agency Approvals
  CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993

 

Mechanical & General Specifications
Size
  136 x 245 x 63 mm (5.4¡° x 9.6 x 2.5)
Weight
  1.2 kg (2.7 lb)
Batteries
Internal Battery Pack Li-ion: 7.2V, 4400mAh, 30 Wh
Battery Life
  >8 hours typical
Battery Replacement
  Replace without opening calibrator; no tools required
Side port connections
  Pressure module connector
  USB Connector to interface to your PC
  Digital instrument (HART) connector
  Connection for optional battery charger/eliminator
Data storage capacity
  1 week of calibration procedures results
90 day specifications
  The standard specification interval for the 750 Series are 1 and 2 years.
  Typical 90 day measurement and source accuracy can be estimated by
  dividing the one year "% of reading" or "%of output" specifications by 2.
  Floor specifications, expressed as "% of full scale" or
  "counts" or "ohms" remain constant.

 

Temperature, Resistance Temperature Detectors
Degrees or % of reading
Type (¥á)Range ¡ÆCMeasure ¡ÆC1Source currentSource ¡ÆCAllowable current2
  1 year2 year  1 year2 year 
100 ¥Ø Pt (385)-200 to 100
100 to 800
0.07 ¡ÆC
0.02 % + 0.05 ¡ÆC
0.14 ¡ÆC
0.04 % + 0.10 ¡ÆC
1 mA 0.05 ¡ÆC
0.0125 % + 0.04 ¡ÆC
0.10 ¡ÆC
0.025 % + 0.08 ¡ÆC
0.1 mA to 10 mA
200 ¥Ø Pt (385)-200 to 100
100 to 630
0.07 ¡ÆC
0.02 % + 0.05 ¡ÆC
0.14 ¡ÆC
0.04 % + 0.10 ¡ÆC
500 ¥ìA 0.06 ¡ÆC
0.017 % + 0.05 ¡ÆC
0.12 ¡ÆC
0.034 % + 0.10 ¡ÆC
0.1 mA to 1 mA
500 ¥Ø Pt (385)-200 to 100
100 to 630
0.07 ¡ÆC
0.02 % + 0.05 ¡ÆC
0.14 ¡ÆC
0.04 % + 0.10 ¡ÆC
250 ¥ìA0.06 ¡ÆC
0.017% + 0.05 ¡ÆC
0.12 ¡ÆC
0.034 % + 0.10 ¡ÆC
0.1 mA to 1 mA
1000 ¥Ø Pt (385)-200 to 100
100 to 630
0.07 ¡ÆC
0.02 % + 0.05 ¡ÆC
0.14 ¡ÆC
0.04% + 0.10 ¡ÆC
150 ¥ìA 0.06 C
0.017 % + 0.05 ¡ÆC
0.12 C
0.034 % + 0.10 ¡ÆC
0.1 mA to 1 mA
100 ¥Ø Pt (3916)-200 to 100
100 to 630
0.07 ¡ÆC
0.02 % +0.05 ¡ÆC
0.14 ¡ÆC
0.04 % +0.10 ¡ÆC
1 mA 0.05 ¡ÆC
0.0125 % + 0.04 ¡ÆC
0.10 ¡ÆC
0.025 % + 0.08 ¡ÆC
0.1 mA to 10 mA
100 ¥Ø Pt (3926)-200 to 100
100 to 630
0.08 ¡ÆC
0.02 % +0.06 ¡ÆC
0.16 ¡ÆC
0.04 % +0.12 ¡ÆC
1 mA 0.05 ¡ÆC
0.0125 % + 0.04 ¡ÆC
0.10 ¡ÆC
0.025 % + 0.08 ¡ÆC
0.1 mA to 10 mA
10 ¥Ø Cu (427)-100 to 260 0.2 ¡ÆC 0.4 ¡ÆC 3 mA 0.2 ¡ÆC 0.4 ¡ÆC 0.1 mA to 10 mA
120 ¥Ø Ni (672)-80 to 260 0.1 ¡ÆC 0.2 ¡ÆC 1 mA 0.04 ¡ÆC 0.08 ¡ÆC 0.1 mA to 10 mA

1For two and three-wire RTD measurements, add 0.4¡ÆC to the specifications.
2Supports pulsed transmitters and PLCs with pulse times as short as 1 ms
 

 

 

Temperature, Thermocouples
Type Source ¡ÆC Measure ¡ÆC Source ¡ÆC
  1 year2 years1 year2 years
E-250 to -2001.32.00.60.9
 -200 to -1000.50.80.30.4
 -100 to 6000.30.40.30.4
 600 to 10000.40.60.20.3
N-200 to -1001.01.50.60.9
 -100 to 9000.50.80.50.8
 900 to 13000.60.90.30.4
J-210 to -1000.60.90.30.4
 -100 to 8000.30.40.20.3
 800 to 12000.50.80.30.3
K-200 to -1000.71.00.40.6
 -100 to 4000.30.40.30.4
 400 to 12000.50.80.30.4
 1200 to 13720.71.00.30.4
T -250 to -2001.72.50.91.4
 -200 to 00.60.90.40.6
 0 to 4000.30.40.30.4
B600 to 8001.32.01.01.5
 800 to 10001.01.50.81.2
 1000 to 18200.91.30.81.2
R-20 to 02.32.81.21.8
 0 to 1001.52.21.11.7
 100 to 17671.01.50.91.4
S-20 to 02.32.81.21.8
 0 to 2001.52.11.11.7
 200 to 14000.91.40.91.4
 1400 to 17671.11.71.01.5
C0 to 8000.60.90.60.9
 800 to 12000.81.20.71.0
 1200 to 18001.11.60.91.4
 1800 to 23162.03.01.32.0
L-200 to -1000.60.90.30.4
 -100 to 8000.30.40.20.3
 800 to 9000.50.80.20.3
U-200 to 00.60.90.40.6
0 to 6000.30.40.30.4
BP 0 to 10001.01.50.40.6
 1000 to 20001.62.40.60.9
 2000 to 25002.03.00.81.2
XK-200 to 3000.20.30.20.5
300 to 8000.40.60.30.6