Multi-Channel Datalogger(GRAPHTEC/GL840 serles)
  • Á¦Á¶»çGRAPHTEC
  • ¸ðµ¨¸íGL840 serles
  • ÁÖ¿ä±â´É ±â·Ï¿ë
»óÇ°»ó¼¼Á¤º¸



 


 

 

 

 

 

 

 

 

 

 

Three types of input system enable to measure various phenomena

Input system 1 : Multifunction analog input ports

Contains a highly isolated input system which ensures that signals are not corrupted by noise from other channels. The GL840's inputs are suitable for combined measurements from voltage, temperature, humidity, logic, and pulse signals.
The standard configuration has 20 analog input channels. It is expandable to 200 channels by adding optional 20 channels extension terminal base unit (B-566) and input terminal units (B-564 or B-565). 

 

 

GL840 Main unit specifications

ItemDescription
Model numberGL840-MGL840-WV
Number of analog input channels20 channels in standard configuration, Expandable up to 200 channels
Number of analog input terminalsUp to 10 terminals (20 channels / terminal), standard config: 1
Type of analog input terminalMulti-input type, Withstand-voltage type
Port for digital sensor1 port for the sensor/input terminal/adapter of the GL100
External input/
output *1
Input *2Trigger or Sampling (1 channel), Logic/Pulse (4 channels)
Output *3Alarm (4 channels)
Sampling interval10 ms to 1 hour (10ms to 50ms: voltage only) *4, External signal
Time scale of waveform display1 sec. to 24 hour /division
Trigger,
Alarm function
Trigger actionStart or stop capturing data by the trigger
Repeat actionOff, On (auto rearmed)
Trigger sourceStart: Off, Measured signal, Alarm, External, Clock, Week or Time
Stop: Off, Measured signal, Alarm, External, Clock, Week or Time
Condition SettingCombination: OR or AND
Analog signal: Rising (High), Falling (Low), Window-in, Window-out
Logic signal: Pattern (combination of each input signal in high or low)
Pulse (number of count): Rising (High), Falling (Low), Window-in, Window-out
Alarm outputOutputs a signal when alarm condition occurs in the input signal *5
Pulse input
function
Rotation count
(RPM) mode
Counts the number of pulses per sampling interval and converts to rpm
(rotations per minute), Number of pulses for one rotation can be set to
50, 500, 5000, 50k, 500k, 5M, 50M, 500M rpm/F.S. (rpm./Full Scale)
Accumulating
count mode
Accumulates the number of pulses from the start of measurement
50, 500, 5000, 50k, 500k, 5M, 50M, 500M C/F.S. (Counts/Full Scale)
Instant count
mode
Counts the number of pulses per sampling interval
50, 500, 5000, 50k, 500k, 5M, 50M, 500M C/F.S. (Counts/Full Scale)
Calculation
function
Between channelsAddition, Subtraction, Multiplication, and Division for analog input
StatisticalSelect two calculations from Average, Peak, Maximum, Minimum, RMS
Search functionSearch for analog signal levels, values of logic or pulse or alarm point
in captured data
Interface to PCEthernet (10 BASE-T/100 BASE-TX), USB (Hi-speed), WLAN (using B-568 option)
Storage
device
InternalBuilt-in 4GB Flash Memory *6
ExternalOne SD card slot (Support SDHC memory card, up to 32GB) *7
Saved contentsCaptured data, Setting conditions, Screen copy
Capturing modeMode: Normal, Ring, Relay
Ring: Saves most recent data (Number of capturing data: 1000 to 2000000 points) *8
Relay: Saves data to multiple files without losing data until dada capturing is stopped
Replay dataReplays captured data that was saved in the GL840 (in GBD or CSV format)
Scaling (Engineering unit) functionMeasured value can be converted to specified engineering unit
• Analog voltage: Converts using four reference points (gain, offset)
• Temperature: Converts using two reference points (offset)
• Pulse count: Converts using two reference points (gain)
Action during data capture• Displaying past data (using dual display mode (Current + Past data))
• Hot-swapping the SD memory card
• Saving data in between cursors
DisplaySize7-inch TFT color LCD (WVGA: 800 x 480 dots)
LanguageEnglish, French, German, Chinese, Korean, Russian, Spanish, Japanese
Information *9Waveform in Y-T with digital values, Waveform only, Digital value, Digital values
and statistics values
Operating environment0 to 45 ¨¬C, 5 to 85 % RH (non condensed)
(When operating with battery pack 0 to 40 ¨¬C, charging battery 15 to 35 ¨¬C)
Power sourceAC adapter100 to 240 V AC, 50/60 Hz (1 pc of adapter is attached as standard accessory)
DC power8.5 to 24 V DC (DC drive cable (option B-514) is required)
Battery packMountable two battery packs (battery pack (option B-569): 7.2V DC, 2900mAh)
Power consumption *10Max. 38 VA
External dimensions (W x D x H
in mm, Excluding projections)
Approx. 240 x 158 x 52.5Approx. 240 x 166 x 52.5
Weight *11Approx. 1010 gApprox. 1035 g
*1
Input/Output cable for GL (option B-513) is required to connect the signal.
*2
Input signal;
  • Voltage range: Up to 24V (common ground)
  • Signal type: Voltage, Open collector, Contact (relay)
  • Threshold: Approx. + 2.5 V (Hysteresis: Approx. 0.5V (2.5V to 3V))
*3
Output signal: Open collector (pull-up to 5V by 10k¥Ø resistor)

  • Voltage: Max. 30V,
  • Current: Max. 0.5A,
  • Collector dissipation: Max. 0.2W

*4
Minimum interval varies by number of channels used.
*5
Output port can be specified in each input channel.
*6
The built-in Flash memory is available for units with serial numbers C604XXXX or later. Please contact your local representative for more information.
*7
SD memory card cannot be used on the SD card slot while the wireless LAN unit (opton B-568) is used.
*8
Size of the capture data will be limited to 1/3 of available memory.
*9
Display mode is switched every time the dedicated key is pressed. In magnified digital value mode, the
displayed channel number can be specified. In the waveform display mode, the changing of the time scale will
be effective from the point of the next displayed data.
*10
Rating under maximum power consumption using the AC adapter, with LCD display on, and battery pack(s)
being charged.
*11
Excludes AC adapter and battery pack.  

 


 

 

GL840 Analog input specifications

ItemDescription
Model numberGL840-M, Input terminal B-564GL840-WV, Input terminal B-565
Input methodAll channels isolated balanced input *12, Scans channels for sampling
Type of input terminalScrew terminal (M3 screw)
Measurement
range
Voltage20, 50, 100, 200, 500 mV, 1, 2, 5, 10, 20, 50, 100 V, and 1-5V F.S. (Full Scale)
ThermocoupleType: K, J, E, T, R, S, B, N, W (WRe5-26)
Range: 100, 500, 2000 ¨¬C *13
RTD (Resistance
Temperature Detector)
Type: Pt100, Pt1000 (IEC751), JPt100 (JIS)
Range: 100, 500, 2000 ¨¬C *13
Humidity0 to 100 % RH - using the humidity sensor (option B-530)
FilterOff, 2, 5, 10, 20, 40 (moving average in selected number)
Measurement accuracy *14
Voltage¡¾ 0.1% of F.S. (Full Scale)¡¾ (0.05% of F.S. + 10¥ìV)
Temperature (Thermocouple) *15
 TypeMeasurement range
(TS: Temp Sense)
Measurement accuracyMeasurement accuracy
R0 ¡Â TS ¡Â 100 ¨¬C¡¾ 5.2 ¨¬C¡¾ 4.5 ¨¬C
100 < TS ¡Â 300 ¨¬C¡¾ 3.0 ¨¬C¡¾ 3.0 ¨¬C
300 < TS ¡Â 1600 ¨¬C¡¾ (0.05% of rdg. + 2.0 ¨¬C)¡¾ 2.2 ¨¬C
S0 ¡Â TS ¡Â 100 ¨¬C¡¾ 5.2 ¨¬C¡¾ 4.5 ¨¬C
100 < TS ¡Â 300 ¨¬C¡¾ 3.0 ¨¬C¡¾ 3.0 ¨¬C
300 < TS ¡Â 1760 ¨¬C¡¾ (0.05% of rdg. + 2.0 ¨¬C)¡¾ 2.2 ¨¬C
B400 ¡Â TS ¡Â 600 ¨¬C¡¾ 3.5 ¨¬C¡¾ 3.5 ¨¬C
600 < TS ¡Â 1820 ¨¬C¡¾ (0.05% of rdg. + 2.0 ¨¬C)¡¾ 2.5 ¨¬C
K-200 ¡Â TS ¡Â -100 ¨¬C¡¾ (0.05% of rdg. + 2.0 ¨¬C)¡¾ 1.5 ¨¬C
-100 < TS ¡Â 1370 ¨¬C¡¾ (0.05% of rdg. + 1.0 ¨¬C)¡¾ 0.8 ¨¬C
E-200 ¡Â TS ¡Â -100 ¨¬C¡¾ (0.05% of rdg. + 2.0 ¨¬C)¡¾ 1.0 ¨¬C
-100 < TS ¡Â 800 ¨¬C¡¾ (0.05% of rdg. + 1.0 ¨¬C)¡¾ 0.8 ¨¬C
T-200 ¡Â TS ¡Â -100 ¨¬C¡¾ (0.1% of rdg. + 1.5 ¨¬C)¡¾ 1.5 ¨¬C
-100 < TS ¡Â 400 ¨¬C¡¾ (0.1% of rdg. + 0.5 ¨¬C)¡¾ 0.6 ¨¬C
J-200 ¡Â TS ¡Â -100 ¨¬C¡¾ 2.7 ¨¬C¡¾ 1.0 ¨¬C
-100 < TS ¡Â 100 ¨¬C¡¾ 1.7 ¨¬C¡¾ 0.8 ¨¬C
100 < TS ¡Â 1100 ¨¬C¡¾ (0.05% of rdg. + 1.0 ¨¬C)¡¾ 0.6 ¨¬C
N-200 ¡Â TS < 0 ¨¬C¡¾ (0.1% of rdg. + 2.0 ¨¬C)¡¾ 2.2 ¨¬C
0 ¡Â TS ¡Â 1300 ¨¬C¡¾ (0.1% of rdg. + 1.0 ¨¬C)¡¾ 1.0 ¨¬C
W0 ¡Â TS ¡Â 2000 ¨¬C¡¾ (0.1% of rdg. + 1.5 ¨¬C)¡¾ 1.8 ¨¬C
R.J.C.¡¾ 0.5 ¨¬C¡¾ 0.3 ¨¬C
Temperature (RTD) *16
 TypeMeasurement range
(TS: Temp Sense)
AccuracyAccuracy
Pt100-200 ¡Â TS ¡Â 100 ¨¬C¡¾ 1.0 ¨¬C¡¾ 0.6 ¨¬C
100 < TS ¡Â 500 ¨¬C¡¾ 0.8 ¨¬C
500 < TS ¡Â 850 ¨¬C¡¾ 1.0 ¨¬C
Pt1000-200 ¡Â TS ¡Â 100 ¨¬C¡¾ 0.8 ¨¬C¡¾ 0.6 ¨¬C
100 < TS ¡Â 500 ¨¬C¡¾ 0.8 ¨¬C
JPt100-200 ¡Â TS ¡Â 100 ¨¬C¡¾ 0.8 ¨¬C¡¾ 0.6 ¨¬C
100 < TS ¡Â 500 ¨¬C¡¾ 0.8 ¨¬C
A/D converterSigma-Delta type, 16 bits (effective resolution: 1/40000 of the measuring full range)
Input resistance1M¥Ø ¡¾5%
Allowable signal source resistanceUp to 300¥ØUp to 100¥Ø
Maximum
input voltage
Between
(+) / (-) terminal
20 mV to 2 V range: 60 Vp-p,
5 V to 100 V range: 110 Vp-p
Channels ((-) / (-))60 Vp-p600 Vp-p
Channel / GND60 Vp-p300 Vp-p
Max. voltage
(withstand)
Between channels350 Vp-p (1 minute)600 Vp-p
Channel / GND350 Vp-p (1 minute)2300 Vrms AC (1 minute)
*12
The terminal "b" for using the RTD is connected each other across all channels.
*13
If the specifications of the temperature sensor is lesser or greater than the selected measurement range, GL840
can measure up to the specifications of the sensor.
*14
Subject to the following conditions:
  • Room temperature is 23 oC ¡¾ 5 oC.
  • When 30 minutes or more have elapsed after power has turned on.
  • Filter is set to 10.
  • Sampling rate is set to 1 sec, using 20-channel in GL840-M and 10-channel in GL840-WV.
  • GND terminal is connected to ground.
*15
Wire size of thermocouple used is 0.32mm diameter in the T or K type and 0.65mm diameter in other types.
*16
Supports 3-wire type sensor.