AC Clamp Adaptor(KYORITSU/8112), 120A
  • Á¦Á¶»çKYORITSU
  • ¸ðµ¨¸í8112, 8112BNC
  • ÁÖ¿ä±â´É ±³·ù Àü·ù ÃøÁ¤¿ë
»óÇ°»ó¼¼Á¤º¸



 

 

Clamp Sensors/Adaptors

MODEL 8112

• AC CLAMP ADAPTOR
• Model 8112 clamp adaptor is designed to be an AC current/voltage conversion probe capable of measuring AC current from 0.1mA to 120A in conjunction with digital multimeters.
Simple to operate, the unit has range switch position at 200mA,2A and 20A. It also features safety design throughout with no exposed metal parts.
  • ¥õ8 MAX120A ACA
RangeMeasuring RangesOutput VoltageAccuracyFrequency Response
200mAAC 0 - 500mAAC1V/A
(1000mA¡æ1V)

¡¾1.5%rdg¡¾0.2mA50Hz - 1kHz
AC 0 - 1000mA¡¾3%rdg¡¾0.4mA40Hz - 10kHz
2AAC 0 - 20AAC100mV/A
(20A¡æ2V)

¡¾1%rdg¡¾1mA40Hz - 1kHz
¡¾1.5%rdg¡¾2mA1k - 10kHz
20A

AC 0 - 20AAC10mV/A
(120A¡æ1.2V)

¡¾1%rdg¡¾0.01A40Hz - 1kHz
AC 20 - 60A¡¾2.5%rdg50Hz - 10kHz
AC 60 - 120A¡¾2.5%rdg100Hz - 10kHz
Conductor Size¨ª8mm max.
Frequency Characteristics30Hz - 100kHz (-3dB)
Safety StandardIEC 61010-1 CAT II 100V Pollution Degree 2
Withstand Voltage500V AC for 1 minute
Dimensions153(L) ¡¿ 18(W) ¡¿ 23(D)mm
Weight100g approx.
Accessories9057 (Carrying Case)
Instruction Manual

 

 

 

 

Clamp Sensors/Adaptors

MODEL 8112BNC

• AC Clamp Adaptor
• Model 8112BNC is an AC clamp adaptor designed for use with oscilloscopes. Output cord has a BNC connector which enables direct observation of current waveform on oscilloscope. Refer to a diagram shown at left for application. Specifications are same as those for Model 8112.
  • ¥õ8 MAX120A ACA
Example of Model 8112 BNC application

RangeMeasuring RangesOutput VoltageAccuracyFrequency Response
200mAAC 0 - 500mAAC1V/A
(1000mA¡æ1V)

¡¾1.5%rdg¡¾0.2mA50Hz - 1kHz
AC 0 - 1000mA¡¾3%rdg¡¾0.4mA40Hz - 10kHz
2AAC 0 - 20AAC100mV/A
(20A¡æ2V)

¡¾1%rdg¡¾1mA40Hz - 1kHz
¡¾1.5%rdg¡¾2mA1k - 10kHz
20A

AC 0 - 20AAC10mV/A
(120A¡æ1.2V)

¡¾1%rdg¡¾0.01A40Hz - 1kHz
AC 20 - 60A¡¾2.5%rdg50Hz - 10kHz
AC 60 - 120A¡¾2.5%rdg100Hz - 10kHz
Conductor Size¨ª8mm max.
Frequency Characteristics30Hz - 100kHz (-3dB)
Safety StandardIEC 61010-1 CAT II 100V Pollution Degree 2
Withstand Voltage500V AC for 1 minute
Dimensions153(L) ¡¿ 18(W) ¡¿ 23(D)mm
Weight100g approx.
Accessories9057 (Carrying Case)
Instruction Manual